Current and voltage sensors: shunts, Hall sensors and CTs

Shunts, Hall-effect sensors, current transformers and Rogowski coils measure current; dividers scale voltage. How each works and how each misleads you.

Core

Four ways to sense current, and one way to scale voltage A shunt resistor, a Hall-effect sensor, a current transformer and a Rogowski coil, each with what its output does and how it can mislead you, and a resistor divider feeding an isolated amplifier for voltage. Shunt resistor 4-wire Small voltage across it bigger shunt: bigger signal, but more heat (P = I²R) Watch for resistance changes as it heats; copper: about 3 900 ppm/°C Check heat against the derating curve Hall-effect sensor Senses the magnetic field isolated from the conductor; works for DC and AC Watch for offset, temperature drift, noise: higher than a shunt's (one maker); outside magnetic fields affect it Current transformer (CT) Needs a changing field, so it cannot measure DC Watch for Secondary open while current flows: a very high voltage, a safety hazard one maker's warning; a US work rule bars it Rogowski coil Output follows di/dt 90° out of phase, so an integrator (circuit or software) rebuilds it Watch for no core to saturate, but it picks up stray fields from nearby wires Used in electricity meters Voltage High voltage e.g. 800 V Resistor divider as many as 15 resistors 2 V Isolated amplifier upper and lower resistors of different types can drift apart
Four ways to sense current, with what each one's output does and how it can mislead you, and a resistor divider scaling a high voltage for an isolated amplifier.

Key points

  1. A shunt is a low-value resistor in the current path; the current makes a small voltage across it, read by an amplifier and ADC. A larger shunt gives a bigger signal, so amplifier offset matters less, but wastes more power as heat (P = I²R). One maker says shunts are typically preferred up to 100 A on rails below 100 V.123
  2. A shunt's resistance changes as it heats, adding error; copper's changes about 3 900 ppm/°C, so copper leads add more. A 4-wire (Kelvin) connection senses across the shunt alone, fixing layout, not the shunt's drift. One maker says to check terminal temperature at maximum nominal operation against the derating curve.42567
  3. A Hall-effect current sensor senses the conductor's magnetic field with no electrical connection, so it is inherently isolated, and works for DC and AC. One maker rates its offset, drift, noise and pickup of outside fields worse than a shunt's; one Hall sensor's datasheet names nearby magnets and high-current traces.83910
  4. A current transformer (CT) needs a changing magnetic field, so it cannot measure DC. One maker's guide says its secondary must never be left open-circuit while primary current flows: a very high voltage results, a significant safety hazard. A US rule for power-line construction bars opening an energised CT's secondary.811
  5. One maker lists Rogowski coils, CTs and shunts as the commonest electricity-meter current sensors. The coil is air-cored; its output follows di/dt, 90° out of phase, so it reads only changing current and an integrator (circuit or software) rebuilds it. With no core it cannot saturate, but picks up stray fields.4812
  6. In one maker's example, an 800 V bus is scaled by a resistor divider to an isolated amplifier's 2 V input range. Another of its notes says makers limit each resistor's voltage, so a divider can need as many as 15, and upper and lower resistors of different types can drift apart with temperature, compounding over time.1314
  7. India's CEA metering regulations (2006, amended to 2026) set accuracy classes: consumer meters to 650 V, 1.0 direct-connected or 0.5S with CTs; 0.5S above 650 V to 33 kV; 0.2S above 33 kV and for interface meters, bar 1.0/0.5S smart meters for open access to 650 V. CTs and VTs must be no less accurate than their meter.151617as of 19 November 2026

Common mistakes

  • Leaving a current transformer's secondary open while current flows in the primary. One maker warns that a very high voltage results, a significant safety hazard.
  • Expecting a CT or a Rogowski coil to read DC. A CT needs a changing magnetic field and a Rogowski coil follows the rate of change of current; a Hall-effect sensor works for DC.
  • Choosing a bigger shunt for a bigger signal without checking the heat. Power rises with the square of current, and a shunt's resistance changes as it heats.

Sources

  1. Integrating the Current Sensing Signal Path (SBOA167B) · Texas Instruments · Reference
  2. Guide for PCB Layout of Sensing Lines Using Shunt Resistor (68AN100E) · ROHM Co., Ltd. · Reference
  3. Comparing Parameters of Shunt-Based and Hall-Based Current Sensors (AN296301) · Allegro MicroSystems · Reference
  4. Benefits of Switching to Rogowski Coil in E-metering (SBAA762) · Texas Instruments · Reference
  5. Copper Wire Tables (Handbook 100) · US National Bureau of Standards (now NIST) · Reference
  6. Low Temperature Drift Integrated Shunt vs Active Temperature Compensation of Shunt (SLYA032) · Texas Instruments · Reference
  7. Shunt Resistor Selection for Isolated Data Converters (SBAA460) · Texas Instruments · Reference
  8. High Current Measurement Application Guide · Yokogawa Test & Measurement · Reference
  9. Comparing Shunt- and Hall-Based Isolated Current-Sensing Solutions in HEV/EV (SBAA293C) · Texas Instruments · Reference
  10. TMCS1100 Hall-Effect Current Sensor data sheet (SBOS820C) · Texas Instruments · Datasheet
  11. 29 CFR 1926.967 Special conditions (Subpart V, Electric Power Transmission and Distribution) · US Occupational Safety and Health Administration · Regulation
  12. Understanding the Design Challenges of Rogowski Coils (SDAA566) · Texas Instruments · Reference
  13. High-Accuracy Isolated Voltage Measurements in HEV/EV Subsystems Using AMC1311-Q1 and AMC1211-Q1 (SBAA289) · Texas Instruments · Reference
  14. Increased Accuracy and Performance with Integrated High Voltage Resistor Isolated Amplifiers and Modulators (SBAA646A) · Texas Instruments · Reference
  15. Central Electricity Authority (Installation and Operation of Meters) (Amendment) Regulations, 2019 · Central Electricity Authority, Government of India · Regulation
  16. Central Electricity Authority (Installation and Operation of Meters) Amendment Regulations, 2026, with corrigendum · Central Electricity Authority, Government of India · Regulation
  17. Central Electricity Authority (Installation and Operation of Meters) Regulations, 2006 · Central Electricity Authority, Government of India · Regulation