Pressure sensors: gauge, absolute and differential

Gauge, absolute, differential and sealed pressure sensors, how a piezoresistive MEMS die works, and how temperature, weather and overpressure mislead you.

Core

Pressure references, ratings and what moves the reading Four pressure references shown as a diaphragm with process pressure on one side and a different reference on the other; below, one part's overpressure and burst ratings, and the temperature, weather and fault effects that move a pressure reading. What the diaphragm compares against Gauge Process Local air Relative to local atmospheric pressure Absolute Process Vacuum Relative to a perfect vacuum Differential Port 1 Port 2 The difference between two ports Sealed gauge Process Sealed reference No vent; reads 0 only when the air matches Ratings of one 0 to 1 psi part 1 psi 30 psi 50 psi Overpressure, 30 psi up to it: returns to specification; beyond it: damage may be permanent Burst, 50 psi past it, the media may escape Not to scale What moves the reading Temperature shifts zero and sensitivity: read the TEB. One part: ±0.25 % at 25 °C, ±1.5 % TEB Weather moves absolute and altitude readings. Absolute, open 5 m tank: 4.7 m to 5.3 m Faults mostly shift the zero: auto-zero at a known pressure
Four pressure references, the overpressure and burst ratings of one part, and the temperature, weather and fault effects that move a pressure reading.

Key points

  1. Gauge pressure is measured relative to local atmospheric pressure, absolute pressure relative to a perfect vacuum, and differential pressure as the difference between two ports. A sealed gauge sensor has a reference pressure sealed inside: it needs no vent, but reads 0 only when the air matches that reference.1234
  2. A piezoresistive MEMS pressure die is silicon etched to a thin diaphragm, stressed when the pressures on its two sides differ. Usually four piezoresistors in it form a Wheatstone bridge that gives a small voltage. One maker lists such sensors for CPAP machines, ventilators and clogged HVAC filters.56789
  3. A piezoresistive die's zero and sensitivity both change with temperature, even after room-temperature calibration. Read the total error band (TEB), the worst error over the full compensated temperature and pressure range. One datasheet: ±0.25 % of span 'accuracy' at 25 °C, ±1.5 % TEB at 0 to 50 °C.6728
  4. One maker defines overpressure (proof pressure) as the most a sensor can take and still meet its specification once pressure is back in range; more may cause permanent damage. Past burst pressure, media may escape. Its 0 to 1 psi gauge part allows 30 psi and 50 psi; 'dry gases only' versions must not see condensation.28
  5. Choose the reference to suit the job. One maker notes that weather can move air pressure by ±30 mbar, so an absolute sensor on an open tank holding 5 m of water could read 4.7 m to 5.3 m. At up to 6 000 bar in water-jet cutting, ±230 mbar is only about 0.004 %, and a non-vented sensor has no vent to let water in.41
  6. A barometric altimeter turns absolute pressure into altitude using a standard atmosphere whose sea-level pressure is 101.325 kPa (1013.25 hPa). Real sea-level pressure changes with the weather, which shows up as altitude error. Setting the local sea-level pressure, or measuring height from a start point, reduces it.101112
  7. One maker says offset is by far the largest remaining error in pressure sensors, and that faults tend to shift the zero; slope changes are very rare. It recommends auto-zero: read the output at a known pressure, such as the port vented to air, and subtract it. Absolute sensors need a vacuum or a barometer for this.1314
  8. In India, the Legal Metrology (General) Amendment Rules, 2026 (G.S.R. 10(E), 7 January 2026) cap the error of a non-invasive, non-automated blood pressure gauge's cuff pressure reading at ±0.4 kPa (±3 mmHg) anywhere on its scale at 15 to 25 °C, and require a range of 0 to at least 260 mmHg.15as of 19 November 2026

Common mistakes

  • Choosing a pressure sensor by its 'accuracy' at 25 °C. The total error band covers the whole compensated temperature range and, in one datasheet, is six times larger.
  • Using an absolute sensor where a gauge reading is wanted, such as the level in an open tank, so changes in the weather move the reading.
  • Trusting a barometric altitude without setting the local sea-level pressure, so a change in the weather looks like a change in height.

Where to go next

Sources

  1. Use of Submersible Pressure Transducers in Water-Resources Investigations (Techniques of Water-Resources Investigations 8-A3, 2004) · U.S. Geological Survey · Reference
  2. Technical Note: Pressure Sensor Glossary of Terms (008200-2-EN, July 2011) · Honeywell Sensing and Control · Reference
  3. Gauge, Absolute, Sealed & Differential Pressure Defined · NOSHOK, Inc. · Vendor page
  4. Gauge and absolute pressure measurement: Absolutely relative (technical article) · WIKA Alexander Wiegand SE & Co. KG · Reference
  5. Structural Engineering in Piezoresistive Micropressure Sensors: A Focused Review (Micromachines 14(8):1507, 2023) · MDPI, via PubMed Central (open access) · Reference
  6. Advances in high-performance MEMS pressure sensors: design, fabrication, and packaging (Microsystems & Nanoengineering 9:156, 2023) · Springer Nature (open access, CC BY 4.0) · Reference
  7. The High Cost of Low Cost Pressure Sensors (white paper, 008317-2-EN, June 2017) · Honeywell · Reference
  8. Basic Board Mount Pressure Sensors, ABP Series datasheet (32305128, Issue 2) · Honeywell · Datasheet
  9. Sensor and Switch Solutions for HVAC Applications (000815-8-EN, October 2025) · Honeywell · Vendor page
  10. AN4528: Pressure Altimetry using the MPL3115A2, Rev. 0 (July 2012) · NXP Semiconductors (originally Freescale Semiconductor) · Reference
  11. JetStream: Air Pressure · NOAA National Weather Service · Reference
  12. Pilot's Handbook of Aeronautical Knowledge (FAA-H-8083-25C), Chapter 8: Flight Instruments · Federal Aviation Administration · Reference
  13. Auto-Zero Calibration Technique for Pressure Sensors (technical note, 2021) · Honeywell · Reference
  14. Pressure or Airflow Sensors? (technical note, 008329-2-EN, April 2021) · Honeywell · Reference
  15. Legal Metrology (General) Amendment Rules, 2026, G.S.R. 10(E), 7 January 2026 (Eighth Schedule, Part VII-A: Non-invasive non-automated sphygmomanometers) · Department of Consumer Affairs, Ministry of Consumer Affairs, Food and Public Distribution · Regulation