Case studies · Client project

A water-monitoring sub-system for a robotics startup

Four low-cost water sensors, characterised on the bench before anyone trusts a number from them, then integrated on a bench rig.

  • Client project
  • Sensing
  • TRL 4
The integrated bench rig on a table: a black tank under a clear acrylic sheet on wooden posts, with the pH probe, the turbidity sensor and a red level-sensor board at the tank, blue water pipes at the left, a pH signal board with a green LED, and jumper wires running to a controller on a breadboard beside a laptop.
The sensors integrated on a bench rig: the tank and its probes, the controller on the breadboard, and the laptop running the dashboard.

The problem

A robotics startup wanted to know whether low-cost sensors could watch water quality well enough to build on: turbidity, total dissolved solids (TDS), pH and water level.

What we did

We selected four sensor modules and tested each one on an Arduino Uno R4 WiFi, in dry air and in household liquids, before any calibration. The point was to find what would make a reading untrustworthy while it was still cheap to find out.

What we found

  1. pH. Acidic water moved the output by +0.920 V, salt water by only +0.044 V. The probe responds strongly to acidity and only slightly to salt.
  2. Turbidity. One cloudy sample read anywhere across a 0.83 V span, depending only on how recently it was stirred.
  3. TDS. The module ranked RO, drinking and tap water correctly, but read distilled water out of order.
  4. Water level. The switch reported dry and wet correctly, with no chatter.

Each effect became a rule for the calibration that comes next: six rules in all.

Where it stands

Technology readiness level 4. The sensors now work together as one system on a bench rig and send their readings over Bluetooth Low Energy to a live dashboard. Each reading on it says how far it can be trusted: “not calibrated”, “accuracy not assessed”, “not verified in the final tank”. Calibration against certified standards comes next.

One minute with the rig running: the sensors in the tank, the controller, the live dashboard and its event log. No sound.

The full write-up

The bench work was done by Nanupatruni Mahesh Kumar and reviewed by Ravi Kumar Chintalapudi. The whitepaper covers that bench stage (level 3): the method, every reading, and the path to level 4.