# Water quality sensors: turbidity, TDS and pH

Turbidity, TDS and pH sensors measure stand-ins: scattered light, conductivity and a voltage. How each works, how each misleads you, and how to check it.

Page: https://dharanova.com/learn/water-quality-sensors · Track: Sense · Level: Core · Facts last checked 19 November 2026

Read first: [What a sensor actually is](https://dharanova.com/learn/what-is-a-sensor.md), [Sensor families](https://dharanova.com/learn/sensor-families.md), [Accuracy, precision and resolution](https://dharanova.com/learn/accuracy-precision-resolution.md), [Calibration, drift and traceability](https://dharanova.com/learn/calibration-and-drift.md)

Diagram: Turbidity: a lamp shines into the water and light scattered at 90° is measured; EPA 180.1 uses a tungsten lamp and NTU, ISO 7027-1:2016 NTU (equal to FNU), above 800 nm; bubbles and scratches read high, colour and fouling low. TDS: electrodes measure conductivity in µS/cm; times 0.5 to 0.9 gives TDS in mg/L. It rises 0.5 to 3 % per °C, so is corrected to 25 °C. pH: a glass electrode gives 59.16 mV per pH at 25 °C; check with two or more buffers (pH 4.01, 7.00, 10.01) and a slope of 95 to 101 %. Bottom: IS 10500:2012 limits, turbidity 1 / 5 NTU, TDS 500 / 2000 mg/L, pH 6.5 to 8.5.

## Key points

1. A turbidimeter shines light into water and measures light scattered at 90°: more scattered light means higher turbidity. US EPA Method 180.1 uses a tungsten lamp and NTU. ISO 7027-1:2016 reports nephelometric readings in NTU, numerically equal to FNU, and notes that measuring above 800 nm limits colour interference. [1][2]
2. Several things bias turbidity readings. Fine air bubbles and scratched optics or sample tubes read high; dissolved colour and light-absorbing material read low, and so does fouling that blocks the beam. Instruments of different designs commonly disagree on the same natural water, even when calibrated the same way. [1][3]
3. A TDS meter does not weigh dissolved solids: it measures electrical conductivity and converts it with a factor. For natural waters, TDS in mg/L is typically 0.5 to 0.9 times the conductivity in µS/cm. Meters referenced to sodium chloride (NaCl) use about 0.49; those referenced to the '442' salt mix about 0.68. [4][5]
4. Conductivity rises by 0.5 to 3 % per °C, depending on the ions present, so meters also measure temperature and report the value corrected to 25 °C, usually with a linear factor of 0.019 to 0.020 per °C or a nonlinear one. Below pH 4, these standard corrections can give errors as large as 50 %. [5]
5. A glass pH electrode develops a voltage across its glass membrane that follows the Nernst equation: ideally 59.16 mV per pH unit at 25 °C, with 0 mV near pH 7 for most electrodes. The slope changes with temperature, so most pH systems include a temperature sensor and compensate automatically. [6]
6. A pH meter is calibrated with two or more standard buffers that bracket the sample's pH, commonly pH 4.01, 7.00 and 10.01. Electrode performance degrades with use; the USGS treats a slope outside 95 to 101 % of the Nernst value (56.2 to 59.8 mV per pH unit at 25 °C) as a sign to clean or replace the electrode. [6]
7. India's drinking water standard IS 10500:2012 (reaffirmed 2023), to Amendment No. 2 (2018), sets pH at 6.5 to 8.5 with no relaxation. Turbidity is 1 NTU acceptable, 5 NTU permissible; total dissolved solids 500 mg/L acceptable, 2000 mg/L permissible. Permissible limits apply only where no other source exists. (as of 19 November 2026) [7][8][9]

## Common mistakes

- Reading turbidity with fine air bubbles in the sample, for example just after shaking it. Bubbles read as extra turbidity, so remove them before you read.
- Treating a low TDS reading as safe water. The meter only measures conductivity; it says nothing about microbes or contaminants that don't conduct.
- Calibrating a pH probe at one buffer, or with buffers that don't bracket the sample, then trusting readings far from them.

## Where to go next

- [Uncalibrated bench characterisation of low-cost turbidity, TDS, pH and level sensors on Arduino Uno R4 WiFi](https://dharanova.com/whitepapers/uncalibrated-water-sensors-arduino-uno-r4-wifi.md): a whitepaper from our lab

## Sources

1. [Method 180.1: Determination of Turbidity by Nephelometry, Revision 2.0 (August 1993)](https://www.epa.gov/sites/default/files/2015-08/documents/method_180-1_1993.pdf), U.S. Environmental Protection Agency
2. [DIN EN ISO 7027-1:2016-11 Water quality - Determination of turbidity - Part 1: Quantitative methods (ISO 7027-1:2016) (catalogue page)](https://www.din.de/en/wdc-beuth:din21:245621388), DIN
3. [National Field Manual for the Collection of Water-Quality Data, Chapter A6, Section 6.7: Turbidity, version 2.1 (September 2005)](https://pubs.usgs.gov/twri/twri9a6/twri9a67/twri9a_Section6.7_v2.1.pdf), U.S. Geological Survey
4. [Obtaining Accurate Readings With TDS Meters](https://knowledge.reagecon.com/wp-content/uploads/2021/09/Obtaining-Accurate-Readings-with-TDS-Meters.pdf), Reagecon
5. [National Field Manual for the Collection of Water-Quality Data, Chapter A6.3: Specific Conductance (Techniques and Methods 9-A6.3, 2019)](https://pubs.usgs.gov/tm/09/a6.3/tm9-a6_3.pdf), U.S. Geological Survey
6. [National Field Manual for the Collection of Water-Quality Data, Chapter A6.4: Measurement of pH (Techniques and Methods 9-A6.4, 2021)](https://pubs.usgs.gov/tm/09/a6.4/tm9a6.4.pdf), U.S. Geological Survey
7. [IS 10500:2012 Drinking Water: Specification (Second Revision), with Amendments No. 1 (2015) and No. 2 (2018)](https://consumerhelpline.gov.in/public/assets/docs/bis/IS10500-2012.pdf), Bureau of Indian Standards
8. [IS 10500 amendments (BIS e-Sale search result: Amd. 1:2015, Amd. 2:2018, Amd. 3:2021, Amd. 4:2021)](https://standardsbis.bsbedge.com/BIS_Amendments.aspx?parentid=10500_2012_Reaff2023_AMD4&stdno=IS%2010500), Bureau of Indian Standards (standards portal operated by BSB Edge)
9. [Programme of Work, FAD 14: Drinking Water and Carbonated Beverages (generated 19-11-2026)](https://www.services.bis.gov.in/php/BIS_2.0/bisconnect/pow_new/Pow/download_pow_pdf_dept_commtt/61/281), Bureau of Indian Standards