# Calibration, drift and traceability

A calibration describes an instrument at one time and under set conditions. How readings drift, how to check them again, and what traceable means.

Page: https://dharanova.com/learn/calibration-and-drift · Track: Sense · Level: Core · Facts last checked 19 November 2026

Read first: [Accuracy, precision and resolution](https://dharanova.com/learn/accuracy-precision-resolution.md)

Diagram: Top: traceability, a documented, unbroken chain of calibrations from a reference, through two calibrations, to the instrument; a widening band shows uncertainty growing along the chain. Middle, a chart not to scale: after a calibration the reading drifts away from the reference value over time, smoothly and in one step; at the recalibration the gap is the drift found. The recalibration interval is set by risk, then reviewed. Bottom: calibration is not adjustment. Calibrate (relates readings to a reference), adjust (changes the instrument), then recalibrate, usually needed after adjusting.

## Key points

1. Calibration and adjustment are different operations. Calibration establishes how an instrument's readings relate to values from a reference, with their uncertainties. Adjustment changes the instrument so it reads as prescribed. Calibration comes first, and after adjusting, the instrument usually needs recalibrating. [1]
2. Instrumental drift is a continuous or incremental change in an instrument's readings over time, caused by changes in the instrument itself, not by the measured quantity or by influences such as temperature. Recalibration reveals it; the drift found can show that shorter intervals are needed, or longer ones possible. [1][2]
3. A calibration is made under specified conditions. Many load-cell specifications, such as offset and sensitivity (gain), are given at one temperature, typically 25 °C; both change with temperature, and ADCs drift too. Correcting that can need calibration points across the operating temperature range. [1][3]
4. For a bridge sensor such as a load cell, a one-point calibration with no load removes offset but leaves gain error. A two-point calibration adds a calibrated test load to correct gain; 80% or more of the range is typically enough. It assumes a linear response, so non-linearity and drift need more points. [3]
5. Metrological traceability means a result can be related to a reference through a documented, unbroken chain of calibrations, each adding uncertainty, so uncertainty grows along the chain. Traceable is not the same as good enough: it does not ensure the uncertainty suits the purpose, or that no mistakes were made. [1]
6. There is no universal best practice for setting recalibration intervals. ILAC-G24/OIML D 10 bases the initial choice mainly on risk, including drift and wear, use and environment, then reviews it from calibration history and intermediate checks. Fixed intervals are not recommended unless a normative document sets one. (as of 18 November 2026) [2]
7. In India, NABL (National Accreditation Board for Testing and Calibration Laboratories), a constituent board of the Quality Council of India, gives voluntary accreditation to calibration labs under ISO/IEC 17025. It is a mutual recognition signatory to APAC and Global ACI, which took over ILAC's work in 2026. (as of 18 November 2026) [4][5]

## Common mistakes

- Treating a factory or one-time calibration as permanent. Readings drift, so plan when to check against a reference again.
- Calibrating at room temperature, then deploying somewhere much hotter or colder and trusting the same correction.
- Trimming a sensor until it agrees with a second meter of unknown calibration. Agreeing with an instrument whose calibration is unknown is not traceability.

## Where to go next

- [Uncalibrated bench characterisation of low-cost turbidity, TDS, pH and level sensors on Arduino Uno R4 WiFi](https://dharanova.com/whitepapers/uncalibrated-water-sensors-arduino-uno-r4-wifi.md): a whitepaper from our lab
- [Water quality sensors: turbidity, TDS and pH](https://dharanova.com/learn/water-quality-sensors.md): Turbidity, TDS and pH sensors measure stand-ins: scattered light, conductivity and a voltage. How each works, how each misleads you, and how to check it.
- [Air quality sensors: gases, VOCs, CO₂ and particles](https://dharanova.com/learn/air-quality-sensors.md): How low-cost MOX gas, NDIR CO₂ and optical particle sensors work, what their numbers really are, and how each one can mislead you.
- [Temperature and humidity sensors](https://dharanova.com/learn/temperature-and-humidity-sensors.md): How RTDs, thermistors, thermocouples and capacitive RH sensors work, and how heat, sunlight and damp air make their readings wrong.
- [Pressure sensors: gauge, absolute and differential](https://dharanova.com/learn/pressure-sensors.md): Gauge, absolute, differential and sealed pressure sensors, how a piezoresistive MEMS die works, and how temperature, weather and overpressure mislead you.
- [Level and flow sensors](https://dharanova.com/learn/level-and-flow-sensors.md): Float switches, ultrasonic and pressure sensors measure liquid level; turbine, magnetic and ultrasonic meters measure flow. How each misleads you.

## Sources

1. [JCGM 200:2012 International vocabulary of metrology (VIM), 3rd edition](https://www.bipm.org/documents/20126/2071204/JCGM_200_2012.pdf/f0e1ad45-d337-bbeb-53a6-15fe649d0ff1), BIPM
2. [ILAC-G24:2022 / OIML D 10:2022 Guidelines for the determination of recalibration intervals of measuring equipment](https://www.oiml.org/en/files/pdf_d/d010-e22.pdf), OIML and ILAC
3. [A Basic Guide to Bridge Measurements (SBAA532A)](https://www.ti.com/lit/an/sbaa532a/sbaa532a.pdf), Texas Instruments
4. [National Accreditation Board for Testing and Calibration Laboratories (NABL): home page](https://nabl-india.org/), NABL, Quality Council of India
5. [Global Accreditation Cooperation Incorporated replaces IAF and ILAC](https://www.apac-accreditation.org/global-accreditation-cooperation-incorporated-to-replace-iaf-and-ilac/), Asia Pacific Accreditation Cooperation (APAC)